Gamma spectrum unfolding for a NaI monitor of radioactivity in aquatic systems: Experimental evaluations of the minimal detectable activity
Baré, J.; Tondeur, F. (2011). Gamma spectrum unfolding for a NaI monitor of radioactivity in aquatic systems: Experimental evaluations of the minimal detectable activity. Appl. Radiat. Isot. 69(8): 1121-1124. https://dx.doi.org/10.1016/j.apradiso.2010.11.024
In: Applied Radiation and Isotopes. Elsevier: Amsterdam. ISSN 0969-8043; e-ISSN 1872-9800, more
Peer reviewed article  

Available in  Authors 

Author keywords
    Gamma-ray monitor; MCNP simulation; Spectrum unfolding; Minimaldetectable activity

Authors  Top 
  • Baré, J.
  • Tondeur, F.

Abstract
    This paper deals with the experimental evaluation of the minimal detectable activity achievable by unfolding the gamma spectra of a NaI monitor. An aquatic monitor initially developed by the Institut des Radio-Eléments (IRE) is used for the application. Unfolding of the spectra is performed with GRAVEL, a UMG package code, on the basis of a response matrix obtained with MCNP5.1.40. Experimental data have been measured at IRE, in a 20 m3 seawater tank, for known activities of 137Cs mixed with other gamma emitters (40K, 133Ba, 113Sn and 139Ce). Deconvolution allows one to reduce the MDA of 137Cs by an order of magnitude.

All data in the Integrated Marine Information System (IMIS) is subject to the VLIZ privacy policy Top | Authors